The Model 3900 Low Humidity Generator is an extremelyaccurate means of producing known humidity values forcalibrating and verifying humidity instrumentation. Basedon the combined, NIST proven, fundamental “two-pressure”and “two-temperature” principles, this system willautomatically supply a continuous humidified gas stream,within the frost/dew point range of -95.0°C to 10.0°C, fordays or even wee
Traceable to NIST |
Two-Pressure Two-Temperature Principle |
Push Button (Keypad) Control |
Push Button (Keypad) Control |
Description
The Model 3900 Low Humidity Generator is an extremelyaccurate means of producing known humidity values forcalibrating and verifying humidity instrumentation. Basedon the combined, NIST proven, fundamental “two-pressure”and “two-temperature” principles, this system willautomatically supply a continuous humidified gas stream,within the frost/dew point range of -95.0°C to 10.0°C, fordays or even weeks unattended.
Simply apply power to the system, and the 3900 will power-up ready to purge and/or generate. Humidity setpoint valuesare input by the operator from the front panel keypad and arelimited only by the operational range of the 3900 humiditygenerator.
Features
Key Specifications
Frost / Dew Point Range: –95 to +10 °C
Frost / Dew Point Uncertainty: (–95 to –90 °C) 0.9 °C
(–90 to –80 °C) 0.5 °C
(–80 to –70 °C) 0.2 °C
(–70 to 10 °C) 0.1 °C
Parts Per Million Range: 0.05 to 12000 PPMv
Relative Humidity Range: 0.0002 to 50%
Saturation Pressure Range:Ambient to 300 psiA
Saturation Pressure Uncertainty (10-50 psiA): 0.05
Saturation Pressure Uncertainty (50-300 psiA): 0.30
Saturation Temperature Range: –80 to +15 °C
Saturation Temperature Uncertainty: 0.08 °C
Saturation Temperature Heating/Cooling Rate: 2 Minutes Per °C Average
Gas Flow Rate Range: 0.1 to 5 L/min
Gas Flow Rate Resolution: 0.02 L/min
Gas Type: Air or Nitrogen
Gas Pressure Rating (MAWP): 350 psiG
Test Port: 1/4 Inch Swagelok® Tube Fitting (6.35mm)
Gas Supply (External): 350 psiG, 5 L/min