373LHX & 2500 Hygrometer Calibration Service

373LHX & 2500 Hygrometer Calibration Service

Description

Calibration Parameter: Dew point, Frost point, RH%

Features
Calibration Range: Dew point(-20—+70℃)
Frost point(-25—0℃)
RH(10 to 98% @ 0-70℃)
RH(10 to 98% @ 0-70℃)

Calibration Parameter: Dew point, Frost point, RH%


Traceable to: NIST / SI


Calibration Range: Dew point-20+70℃)

Frost point-250℃)

RH10 to 98% @ 0-70℃)


373LHX Dew Point Hygrometer

  • Dual mirror temperature sensors
  • Removable mirror module for calibration in liquid
  • Heated measuring heads and sampling systems
  • High pressures up to 20000 kPa
  • Temperature controlled sampling system
  • Fast response
  • Touch screen full color LCD user interface



Specifications

Measuring Ranges

Frost/Dew Point: -75...95 °C

Temperature: -50...100 °C

Sample pressure: 500...2’000 hPa abs.


Accuracy

Frost/Dew point (for calibrated range): ≤±0.1°C

Temperature: ≤±0.07°C


Reproducibility

Frost/Dew point: ≤±0.05°C

Temperature: ≤±0.04°C


Standard Features

Digital I/O: RS-232

Mirror cooling: 3-stage Peltier thermoelectric

Mirror temperature sensors: Dual Platinum Resistance Thermometer PRT (Pt-100)

Analog outputs: Two analog outputs, user programmable, -10...+10 V and 4...20 mA

Gas connections: 6mm or 1/4’’ Swagelok

Sample gas flow rate: 0...1 l/min

Sample gas circuit: Stainless steel, 316 / 316L, electro polished

Cooling: Refrigeration

Internal pump: Yes


2500 Generator

  • 0.5 %RH Uncertainty
  • Traceable to NIST
  • Self Contained and Mobile
  • Automated Control of User Setpoints
  • 2500 ControLog® Automation Software
  • HumiCalc® with Uncertainty Software
  • Computerized Internal Transducer Calibration
  • Low Noise Air Compressor with Air Dryer



Specifications

Relative Humidity Range:10 to 95 %RH

Relative Humidity Uncertainty: @ Pc Tc 0.5 %RH

Chamber Temperature Range: 0 to 70 ºC

Chamber Temperature Range: (Optional) –10 to 70 ºC

Chamber Temperature Uncertainty: 0.06 ºC

Chamber Pressure Uncertainty: 0.15% FS

Chamber Pressure Range: Ambient

Gas Flow Rate Range: 5 to 20 L/min

Gas Flow Rate Resolution: 0.02 L/min

Gas Flow Rate Uncertainty: 2% FS

Gas Type: Air or Nitrogen

Gas Pressure Rating (MAWP): 175 psiG

Heating/Cooling Rate: 2.5 Minutes Per ºC Average

Chamber Window: 6" x 6" (152 mm x 152 mm)

Air Supply (External): Clean Oil Free Instrument Air @ 175 psiG & 20 L/min


Applications

Used by NIMI, EPRI and academic institute for calibrating hygrometer

Used by dew point instrument manufacturers, humidity sensor and vapor content meter manufacturers for calibrating