Calibration Parameter: Dew point, Frost point, RH%
Calibration Range: Dew point(-20—+55℃) |
Frost point(-20—0℃) |
RH(5 to 95% @ 0-60℃) |
RH(5 to 95% @ 0-60℃) |
Calibration Parameter: Dew point, Frost point, RH%
Traceable to: NIST
Calibration Range: Dew point(-20—+55℃)
Frost point(-20—0℃)
RH(5 to 95% @ 0-60℃)
Advantages of Chilled Mirrors in the Validation of Climatic Chambers
Specifications
473 Dew Point Mirror
Frost/dew point range: -30 to +70 ̊C
Temperature: -50 to +100 ̊C
Temperature with T12: -200 to +250 ̊C
HygroGen2 Generator
Control range max.: 5…95 %rh and 0…60 °c
(2...99 %rh / -5...60 °c with range extension)
Control stability: <0.02 °c, 0.1 %rh
Temperature gradient: <0.05 °c (15...50 °c), <0.1 °c (5...60 °c), ±0.15 °c (0 °c)
Humidity gradient: 0.1 %rh or better
Typical time to set-point: 3 minutes (35 to 80 %rh change) 5 minutes (20 to 30 °c change)
Temperature sensor: pt100 1/3 din class b
Humidity sensor: hygromer
Control probe accuracy: ±0.8 %rh (23 °c ±5) / ±1.5 %rh (0...60 °c) /
±0.1 °c (23 °c ±5) / ±0.2 °c (0...60 °c)
Chamber volume: 2 litres, effective working volume 1.5 litres, ø 110 mm, 145 mm deep
Applications