MDD-SF6 is an environment-friendly equipment, featuring automatic gas mixing, instrument calibration and instrument verification functionality in one unit. The new design of double diagonal side ventilation and integrated installation technology in a closure rack are originated from and authorized by the German manufacturer. The whole unit is composed of calibration device for automatic gas mixer,
High precision, mix 10-8 ~ 10-2 various concentrations of trace gases |
Multi channel selectable, i.e. random combination of trace gas and background gas |
Good stability, repeatability and high linear precision |
Good stability, repeatability and high linear precision |
Description
SF6 Gas Mixing Dynamic Distribution System MDD-SF6 is an environment-friendly equipment, featuring automatic gas mixing, instrument calibration and instrument verification functionality in one unit. The new design of double diagonal side ventilation and integrated installation technology in a closure rack are originated from and authorized by the German manufacturer. The whole unit is composed of calibration device for automatic gas mixer, workbench for instruments under test/calibration and ventilation. The independent ventilation meets the European laboratory fume hood testing standards. The MDD-SF6 has a simple and smooth frame with easy operation. It offers a pollution-free and safe environment-friendly condition for users who carry out instrument calibration and verification in the laboratory as well as Institute of Metrology.
Features
SF6 Gas Mixing Dynamic Distribution System main part - SF6 Gas Mixer
Principle
Using mass flow to mix 10-8 ~ 10-2 various concentrations of trace gases
Functionality
Automatic mix gas: the device is a computer intelligent gas mixer, all the operating functions are completed by the touch screen of the gas mixer, input the name and concentration of the gases that need to be prepared and diluted, by pressing the touch screen a few times, 10 seconds or so a standard gas of correct concentration is ready;
Automatic calibration: according to the nature of different gases and the environment temperature under experiment, automatic correct flow volume;
Mix multiple sets of gas, or one type of gas at the same time;
Choose different background gas for various configuration (N2, SF6, Air... )
Use of gas: mixed standard gas can be used, also can be charged into the cylinder or air bag to off-site use.
SF6 Gas Mixing Dynamic Distribution System main part – Recovery device of used gas and standard gas
Auto control module, audio and visual alarm
Compressor supervision accumulated counter
Over pressure protection
Mains power module (fused)
Standard gas: concentration and equivalent as per customerized
Recover standard gas and used gas separately
Compressor discharge of standard gas recovery: 1.8m3/h
Compressor discharge of used gas recovery: 1.0m3/h
Ultimate pressure after recovery: 5.0MPa
Specification
Flow range, the number of groups, the maximum dilution ratio can be set according to user requirements
Gas distribution principle: mixed method
Flow linear error: less than ±0.3%
Detection range: the volume ratio of 10-8 ~ 10-2
Gas mixer traceability: traceable to national measurement standards
Limiting dilution ratio: 3000:1
Flow control mode: mass flow meter
Channel number: 4
Calibration gas: SF6 (optional N2, Air, etc.)
Gas distribution outlet flow: 200 ~ 2000mL/min (SF6)
Dilute gas channel flow: 40-2000ml/min (SF6)
Gas concentration repeatability: less than ±0.5%
Gas distribution tubing: PTFE, passivated stainless steel
Ventilation volume: 1000 m3/h
Operating voltage: 220V-240V 50Hz, AC
Operating temperature: -10 ℃ ~ 50 ℃
Ambient humidity: no more than 90%RH
Dimension: 1125mm X 700mm x 1950 mm
Instrument weight: about 130kg